Human LAMC2 enzyme-linked immunoassay kit
Safety datasheet
製品概要
製品名
Human LAMC2 enzyme-linked immunoassay kit
製品タイプ
Elisa Kit
検出感度
0.17 pg/ml (50 μl);2.50 pg/ml (10 μl);
| pg/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0294 | 0.0294 | 0.0294 | |
| 13.72 | 0.0333 | 0.0390 | 0.0362 | 0.0068 |
| 41.15 | 0.0455 | 0.0522 | 0.0489 | 0.0195 |
| 123.46 | 0.0964 | 0.0886 | 0.0925 | 0.0631 |
| 370.37 | 0.2276 | 0.2279 | 0.2278 | 0.1984 |
| 1111.11 | 0.6936 | 0.6883 | 0.6910 | 0.6616 |
| 3333.33 | 2.0760 | 2.0380 | 2.0570 | 2.0276 |
| 10000.00 | 3.9370 | 4.0532 | 3.9951 | 3.9657 |
精度
| Intra-Assay Precision | Inter-Assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average (pg/ml) | 150.6 | 785.4 | 2461.7 | 150.3 | 798.8 | 2646.6 |
| Standard Deviation | 7.0 | 21.8 | 103.1 | 7.5 | 14.7 | 96.5 |
| Coefficient of Variation (%) | 4.7 | 2.8 | 4.2 | 5 | 1.8 | 3.6 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
スパイク回収
The spike recovery was evaluated by spiking 3 levels of human LAMC2 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 79% to 118% with an overall mean recovery of 94%.
サンプル値
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | n.d.-47.67 | 18.8 | 21.95 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated in this assay. No medical histories were available for the donors.
実験タイプ
種属反応性
検出方法
所要時間
標準曲線梯度
インキュベーション回数
2
Detectable Sample
サンプル添加量
製品の特徴
保存条件
2-8 ℃で保存
ターゲット
UNIPROT #
機能
同一ターゲット・同一パスウェイ製品