Human IL-1 alpha enzyme-linked immunoassay kit
Safety datasheet
製品概要
製品名
Human IL-1 alpha enzyme-linked immunoassay kit
製品タイプ
Elisa Kit
検出感度
1.37 pg/ml (50 μl);6.71 pg/ml (10 μl);
| pg/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0080 | 0.0077 | 0.0079 | |
| 2.74 | 0.0134 | 0.0135 | 0.0135 | 0.0056 |
| 8.23 | 0.0254 | 0.0229 | 0.0242 | 0.0163 |
| 24.69 | 0.0543 | 0.0559 | 0.0551 | 0.0473 |
| 74.07 | 0.1547 | 0.1531 | 0.1539 | 0.1461 |
| 222.22 | 0.4364 | 0.4292 | 0.4328 | 0.4250 |
| 666.67 | 1.1850 | 1.2130 | 1.1990 | 1.1912 |
| 2000.00 | 2.9320 | 2.9920 | 2.9620 | 2.9542 |
精度
| Intra-Assay Precision | Inter-Assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average (pg/ml) | 38.1 | 186.6 | 595.9 | 40.2 | 194.7 | 593.9 |
| Standard Deviation | 1.9 | 8.6 | 25.6 | 1.7 | 8.6 | 21.6 |
| Coefficient of Variation (%) | 5.1 | 4.6 | 4.3 | 4.2 | 4.4 | 3.6 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
スパイク回収
The spike recovery was evaluated by spiking 3 levels of human IL-1 alpha into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 100% to 120% with an overall mean recovery of 115%.
サンプル値
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | n.d.-28.38 | 73.3 | 7.14 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated in this assay. No medical histories were available for the donors.
実験タイプ
種属反応性
検出方法
所要時間
標準曲線梯度
インキュベーション回数
2
Detectable Sample
サンプル添加量
製品の特徴
保存条件
2-8 ℃で保存
ターゲット
UNIPROT #
機能
同一ターゲット・同一パスウェイ製品