Human APOC3/Apolipoprotein CⅢ enzyme-linked immunoassay kit(one step)
Safety datasheet
製品概要
製品名
Human APOC3/Apolipoprotein CⅢ enzyme-linked immunoassay kit(one step)
製品タイプ
Elisa Kit
検出感度
0.20 ng/ml (10 μl)
| ng/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0048 | 0.0041 | 0.0045 | |
| 1.37 | 0.0055 | 0.0058 | 0.0057 | 0.0012 |
| 4.12 | 0.0093 | 0.0098 | 0.0096 | 0.0051 |
| 12.35 | 0.0178 | 0.0218 | 0.0198 | 0.0154 |
| 37.04 | 0.0527 | 0.0598 | 0.0563 | 0.0518 |
| 111.11 | 0.2111 | 0.2129 | 0.2120 | 0.2076 |
| 333.33 | 0.9572 | 1.0480 | 1.0026 | 0.9982 |
| 1000.00 | 2.8340 | 2.9620 | 2.8980 | 2.8936 |
精度
| Intra-Assay Precision | Inter-Assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average (ng/ml) | 16.0 | 95.3 | 288.4 | 15.0 | 90.4 | 299.6 |
| Standard Deviation | 1.0 | 4.5 | 22.8 | 0.5 | 5.0 | 25.2 |
| Coefficient of Variation (%) | 6.4 | 4.7 | 7.9 | 3.6 | 5.5 | 8.4 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
スパイク回収
The spike recovery was evaluated by spiking 3 levels of Human APOC3/Apolipoprotein CⅢ into health Human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 70% to 118% with an overall mean recovery of 91%.
サンプル値
| Sample Matrix | Sample Evaluated | Range (µg/ml) | Detectable (%) | Mean of Detectable (µg/ml) |
|---|---|---|---|---|
| Serum | 30 | 9.51-215.14 | 100 | 85.02 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated in this assay. No medical histories were available for the donors.
実験タイプ
種属反応性
検出方法
所要時間
標準曲線梯度
インキュベーション回数
2
Detectable Sample
サンプル添加量
製品の特徴
保存条件
2-8 ℃で保存